EasyManua.ls Logo

Olympus OER-Pro - 7.4 Microbiological Surveillance

Olympus OER-Pro
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7 Routine Maintenance
205
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
5. Connect the water supply piping disinfection hose between the
Air/water/instrument channel connector (gray) in the reprocessing basin and
the water supply piping disinfection connector.
Figure 7.24
6. Close the lid.
7. Press the FUNC SEL button on the subcontrol panel to select “DIS LINES”.
Figure 7.25
Washing case mount
Temperature sensor
Air/water/instrument channel connector
Water supply piping disinfection
connector

Table of Contents

Related product manuals