Chapter 3 Inspection Before Use
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
1. Hold the section marked “PULL” of the detergent/alcohol drawer and pull it
out.
Figure 3.4
2. Push the lock lever on the connector of the tube connected to the detergent
tank to detach the tube.
Figure 3.5
If a few drops of detergent drip from the connector, wipe it
with a piece of clean gauze or similar cloth.
3. Remove the detergent tank.
Detergent tank
Lock lever