Chapter 4 Reprocessing Operations
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
2. Press the FUNC START button on the subcontrol panel to start the water
supply. When water supply starts, the TIME/CODE display on the main
control panel shows a [ ] mark spinning as shown below.
Figure 4.29
3. When the water supply completes, the buzzer beeps three times and the
TIME/CODE display shows [10], which indicates 10 minutes. The time
displayed on the main control panel counts down every minute.
4. When the lid is unlocked, make sure that the water supply has stopped, and
then step on the foot pedal to open the lid.
When the lid is opened, the TIME/CODE display on the main
control panel is reset to [10] and the countdown restarts.
5. Check that there are not any air bubbles produced continuously from the
outer surface of the endoscope and/or leak test air tube while angulating
bending section of the endoscope. Check for leaks at least for 30 seconds.