Chapter 4 Reprocessing Operations
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
4. Connect the equipment-side connector of the connecting tube to the
same-colored connector at the rear center of the reprocessing basin by
pushing the tube’s connector until it clicks. Hang the tag along the
reprocessing basin.
Figure 4.33
5. Gently tug on the connecting tube to confirm that it is firmly attached and
that it will not detach during use.
6. Check the following.
The connecting tube is not kinked.
The endoscopes should not touch the lid.
The endoscopes should be located below the pin that
marks the disinfecting liquid level.
The scope ID tags should be located in the specified
positions.
The cover of the washing case should be closed and the
washing case should be located in the specified position.
When reprocessing an endoscope When reprocessing two endoscopes