Chapter 4 Reprocessing Operations
115
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
3. A jet of fluid exits the hole on the connecting tube during reprocessing.
Make sure that the fluid hits the lid (i.e., the air/water supply channels are
not clogged or abnormal).
Figure 4.36
4. Make sure that a jet of water is output from the water supply/circulation
nozzle to the dome of lid (i.e., the cleaning pump is not clogged or
abnormal).
Figure 4.37

Table of Contents