Chapter 6 Other Functions
159
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
After disinfection, rinse is performed three times. Rinse conditions such as
number of rinsing performed and rinsing time cannot be changed. Time and
water amount of each rinse cycle are shown below. Time and water amount
value vary depending on endoscope models, number of endoscopes loaded and
water supply conditions.
Program [1] is preset to the values (cleaning time,
disinfecting time and disinfectant solution heating) and the
settings of this program cannot be altered.
Program [2] and [3] allow the user to extend the cleaning
time.
The disinfectant solution temperature takes about 4 minutes
for each 1C increase.
If you want to change disinfectant solution from ready-to-use
disinfectant solution to concentrated disinfectant solution,
contact Olympus.
1. Set the power switch to ON.
2. Press the PROG button on the main control panel to select the program to
be set.
Figure 6.5
Number of rinses 123
Time 100 s 110 s 130 s
Rinse water volume 12.4 L
3.3 gallons
14.9 L
3.9 gallons
14.9 L
3.9 gallons
Table 6.3