Chapter 6 Other Functions
185
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
3. After pressing the “ ” button for 5 seconds, the function lamp of subcontrol
panel starts to blink for indicating the current setting of “Error Result
Printout”. Make sure the indication of function lamp on the subcontrol panel
and determine the setting of “Error Result Printout” in accordance with the
description of Table 6.12. If you want to change the setting, keep pressing
the “ ” button. The setting will be changed every 5 seconds by rotation.
For “AUTOMATIC PRINTING” and “PRINT ALL/PRINT
TODAY”, the “Error Result Printout” is set to “include” at the
factory.
Panel display of sub-control panel
Setting of Error Result Printout
PRINT
TODAY/ALL
AUTOMATIC
PRINTING
PRINT
LAST
(If the AUTOMATIC
PRINTING is
activated, this setting
will be reflected)
(Fixed)
–
–
(If the AUTOMATIC
PRINTING is
activated, this setting
will be reflected)
–
includes the error printout
–
excludes the error printout
Table 6.12
When both “PRINT LAST” and “PRINT TODAY” lamp blink
When “PRINT LAST” lamp blinks
When “PRINT TODAY” lamp blinks
When “HEAT LCG” lamp blinks