190
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
3. Unlock each gas filter case cover and open it.
Figure 7.3
4. Remove the old gas filter. The gas filter case designed for the lid has a gas
filter adapter inside of it to prevent water droplet penetration.
Figure 7.4
5. Place a new gas filter in each gas filter case.
Figure 7.5