194
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
8. Press the FUNC START button on the subcontrol panel.
Figure 7.9
Figure 7.10
9. When the water flow stops, press the STOP/RESET button to end the air
purge process, and disconnect the two filter tubes by pushing the lock levers
on their connectors.
Figure 7.11
Connector above water filter housing
Connector below water filter housing