Chapter 7 Routine Maintenance
197
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
5. Place the new water filter directly from the bag into the water filter housing
so that O-ring will be positioned upward.
Figure 7.15
6. Rotate the water filter housing in the direction shown to secure temporarily.
Figure 7.16
To ensure smooth installation, it is recommended to moisten
the O-ring at the head of the water filter with clean water or
ethanol before securing it temporarily.
The rotation drag increases during temporary securing, but
rotate the case all the way until it is stopped.