Chapter 7 Routine Maintenance
207
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
14. Press the FUNC START button on the subcontrol panel again.
Figure 7.28
15. Disconnect the filter tube as soon as disinfectant solution starts to flow from
it continuously.
Figure 7.29
If the temperature of the disinfectant solution is below 20C
(68F), it is heated to 20C (68F). During heating, the
TIME/CODE display blinks without counting down. It restarts
countdown after completion of heating.