Chapter 7 Routine Maintenance
231
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
13. Open the front door and remove the rubber cap from the disinfectant
removal port.
Figure 7.61
14. Put the tube-side end of the disinfectant removal tube in the container
(small), and connect the disinfectant removal tube to the disinfectant
removal port to drain the small amount of residual disinfectant solution from
the disinfectant solution tank (see Figure 7.62).
Figure 7.62
Disinfectant
removal port
Rubber cap