Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
5. Press the FUNC START button on the subcontrol panel. The buzzer beeps,
the disinfectant solution begins draining, and the reprocessing basin is
rinsed. The TIME/CODE display on the main control panel shows a [ ]
mark spinning to indicate that the process is underway.
Figure 7.65
If “DRAIN LCG” is selected and the FUNC START button is
pressed while the lid is open, the buzzer makes several short
beeps, the disinfectant will fill the reprocessing basin, and
error code [E72] is displayed. If error code [E72] is displayed,
take the countermeasure by following the procedure in
Section 8.1, “Troubleshooting guide” on page 274.
6. When draining stops, the long buzzer sounds and the TIME/CODE display
on the main control panel shows [- -] indicating the end of the process.
Figure 7.66

Table of Contents