240
Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
6. Pull out the disinfectant bottle drawer.
Figure 7.72
7. To remove the empty cassette bottles from the disinfectant bottle drawer,
position the bottle mouths upward so that the disinfectant solution does not
come out.
Figure 7.73
8. If any disinfectant solution spills, wipe it away with a clean cloth. If the
disinfectant bottle drawer is dirty, clean using a clean cloth moistened with
neutral detergent solution and then wipe with a clean cloth.
Do not put your hand behind the disinfectant bottle drawer.
Irritation of skin due to contact with concentrated disinfectant
solution, injury by touching a projection, or malfunction of this
equipment may result.
If the cassette cutters are abnormal, do not use the
equipment and contact Olympus. If the cassette cutters are
abnormal, the disinfectant solution cannot be properly
prepared. This will prevent effective endoscope reprocessing
and may cause the equipment to malfunction.
Cassette bottles
Purple cap