Chapter 7 Routine Maintenance
243
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
13. The buzzer generates short beeps and the TIME/CODE display on the main
control panel alternately displays (a) and (b) in Figure 7.78, indicating that
the disinfectant solution is being prepared.
Figure 7.78
• Preparation of disinfectant solution includes single rinse
process to rinse the reprocessing basin, and the required
time for preparation of disinfectant solution is about
5 minutes. While executing the preparation of disinfectant
solution, buzzer beeps intermittently and the main control
panel displays as shown in Figure 7.78. STOP/RESET
button is deactivated during the process.
• A small amount of water comes out of the disinfectant
solution nozzle in the early stage of the process. And a
disinfectant solution is injected from the disinfectant solution
nozzle in the middle stage of the process. These operation
are not a malfunction.
• The required time for preparation of disinfectant solution may
vary depending on the water supply condition.