Chapter 8 Troubleshooting and Repair
277
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
• If the automatic printing function is activated and printout is
not satisfactory due to the error that occurred during the
automatic printing, error code will be displayed on the main
control panel. In this case, print the latest reprocessing
record by “PRINT LAST” to confirm whether the reprocessing
process is completed. Press the “FUNC SEL” button on the
control panel to select “PRINT LAST” and then press the
“START” button. If the printer prints the reprocessing results,
it shows that the previous reprocessing process was
successfully completed. If the printer prints the error results,
it shows that the previous reprocessing process was
interrupted by error indicated in the printed error results.
Error
Code
Problem Possible Causes Remedial Actions
E00 Process is
interrupted.
STOP/RESET
button is pressed
during process.
Wait until automatic processing is finished. During the error
processing, the error code may either blink or light steadily. If
you want to interrupt automatic processing and discharge fluid
inside the equipment, perform air purge. If you want to rinse the
basin, perform rinsing.
E01 Filling the basin
with water takes
too long (water
supply time is
beyond
maximum
setting).
• The water
faucet is not
open enough.
• Water leakage
or clogging in
the water supply
piping.
• Open the water faucet all the way.
• Check if water is leaking from the water supply hose.
• Replace the water filter.
• Clean the mesh filter in the water supply hose connector.
E02 Cleaning fluid is
not discharged.
• Irregularity in
the drain tubing.
• Clogging of the
mesh filter in the
drain port of the
reprocessing
basin.
• Check the drain hose for improper installation.
• Clean the mesh filter in the drain port of the basin.
• Close the lid and perform air purge to remove water.
E04 Cleaning fluid
decreases
during the
cleaning
process.
Internal problem
with the
equipment.
Contact Olympus.
E05 Cleaning fluid
level is too high.
Irregularity in the
fluid level sensor.
Contact Olympus (if water supply will not stop, close the water
faucet).