Chapter 3 Inspection Before Use
61
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
To obtain disinfectant test strips, contact Olympus.
1. Push [PUSH] on the front door to open the front door. Remove the rubber
cap from the disinfectant removal port.
Figure 3.26
2. Push the drain connector into the disinfectant removal port until it clicks.
When connecting the drain connector to the disinfectant
removal port, do not push on the connector’s valve.
Otherwise, disinfectant solution will leak out of it.
3. Place the prepared container below the drain connector, push the
connectors valve, and collect the drained disinfectant solution until it is
deep enough to immerse the reaction area of a test strip.
Disinfectant
removal port
Rubber cap