Chapter 4 Reprocessing Operations
93
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
1. Step on the foot pedal to open the lid.
2. Gently place the control section of the first endoscope in the specified
position (over the white marker) in the retaining rack ensuring that it is below
the index bar (place the control section in the deep part of the reprocessing
basin so that it will not be scratched) (see Figure 4.12).
Figure 4.12
3. Wind the insertion tube clockwise around the retaining rack from the
perimeter in (wind the first turn of the insertion tube outside the black
markings and the second turn inside them) (see Figure 4.13).
Figure 4.13
Setting position
for first control
section
White
marker
Index bar
Control
section
Index bar
Black marking
Black marking