User Calibr
ation
Section 6 User Calibration
6-6
Key Operation Procedure
Perform the operation according to the following procedure.
Moving to Calibration
Level
Parameter Setting Procedure
The parameter character is “cmov.”
The parameter can be changed when the SV display starts to
flash.
If the password is correct, the Unit moves to the calibration level.
If the password is incorrect, the Unit remains in the advanced
function setting level and the next parameter is displayed.
Operation in
Calibration Level
Parameter Operation Procedure
A Move to the advanced function setting level, press the M [MODE]
Key several times and display the “cmov” parameter to move to
the calibration level.
B Press the S [SHIFT] Key to make the SV display flash.
C Use the U [UP] and S [SHIFT] Keys to set the password. The
password is1201”(1201).
D Press the M [MODE] Key to write the password.
Move to the calibration level.
The set value is always 0 after moving
from character display to monitor status.
(Does not move to the calibration level.)
To next parameter
Calibration level
Set value initialization
Password doesn't match.
Password matches.
Change status
cmoV
cmoV
0
0
u
adj
Use the
S
[SHIFT]
and
U
[UP] Keys
to set the password.
30
A Follow the steps outlined above to
move to the calibration level.
The aging timer is displayed.
The aging timer is a 30-minute
countdown timer that counts until 0 is
reached.
A calibration record mark will be
displayed if a user calibration history
exists.
adj
30
Displays “L u.”
L u
30
. adj
Calibration
record mark
L u