SOCT Copernicus PLUS - Service Manual
Turn on the Skantest program and initialize sled + movement.
Warm unit for 40 minutes
When system is warmed perform spectrometer stability test:
In scantest program use option [FAST] to adjust spectrometer.
1. Observe the band height 1 hour from adjusting spectrometer. Signal cannot drop more than
10%. Note: DO NOT USE CALIBRATION OPTIONS DURING TEST.
2. If signal drops more than 10%. Rearrange connecting and placing of CL cable, and repeat test.