SOCT Copernicus PLUS - Service Manual
3. Turn the right screw in accordance with clockwise if the peak is getting wider turn the screw
in the opposite direction ( make turnover of 180
0
) and then turn the screw from the left side
in the same way. Turn alternately until you will get the highest peak ( always turn each of
screw of 180
0
).
4. If the width of peak starts getting smaller after you have turned the screw from the right side
in accordance with clockwise, make the turnover of 180
0
, then turn the screw from the left
side ( also 180
0
). Do it alternately until the peak will be the highest.