SOCT Copernicus PLUS - Service Manual
187
________________________________________________________________________________
Q41: Calibration fail. Please run scantest_lt.
A41: NI driver is not installed.
________________________________________________________________________________
Q42: Unable to start acquisition module.
A42: NI driver is not installed.
________________________________________________________________________________
Q43: The serial read did not complete
with the specified timeout period.
A43: Wrong camera selected in NI drivers.
________________________________________________________________________________
Q44: The serial read did not complete
with the specified timeout period.
A44: Wrong camera selected in NI drivers.
________________________________________________________________________________