SOCT Copernicus PLUS - Service Manual
186
__________________________________________________________________________________
Q37: SOCT_26.exe error.
A37: Synchronization cable disconnected.
CL cable disconnected
(after cable connection device must be restarted).
No voltage on AVIIVA camera
(after cable connection device must be restarted).
________________________________________________________________________________
Q38: Connection with SOCT error. Please Reconnect.
A38: No voltage on AVIIVA camera (during work).
CL cable disconnected (during work).
________________________________________________________________________________
Q39: This application has failed to start because
imaq.dll was not found.
A39: NI driver is not installed.
________________________________________________________________________________
Q40: Unnatural shape of the spectrum.
A40: Wrong camera selected in NI drivers .
________________________________________________________________________________