SOCT Copernicus PLUS - Service Manual
Q34: Noise on tomogram.
A34: Oversaturated spectrum. Adjust gradient filter on reference lane.
__________________________________________________________________________________
Q35: Cannot detect recognizable video source.
A35: CL cable disconnected.
No voltage on AVIIVA camera.
__________________________________________________________________________________
Q36: No tomogram preview.
A36: CL cable disconnected.
No voltage on AVIIVA camera.
Wrong camera is selected in NI drivers.
__________________________________________________________________________________