SOCT Copernicus PLUS - Service Manual
184
Q30: After centering scanners scanner position is wrong.
A30: Broken scanner controller D/A converter - exchange converter
Wrongly attached scanner steering wiring check correctness of connection or exchange
wiring
Broken CPU board 151-506 replace CPU board
Q31: While normal functioning device goes to the “base” position.
A31: Check correctness of attachment of all plugs on PSB board and motor control board
Broken PSB board or Movement control board 151-517 replace it.
Q32: While normal movement of the SOCT head scanners change their status to error status.
A32: Check correctness of the attachment scanners engines connectors, check correctness of
voltage in the main grid, check Switch source power supply.
Q33: When moving to Acquire (Measurement) tab, there is no tomogram visible the Grey
window instead.
A33: Reinstall the National Instruments software.