8 Care and Maintenance
Surveying the Quantum GX2 microCT for Radiation Leakage
Maintenance & Safety Checks
Cleaning the Quantum GX2 microCT on page 81
8.1 Surveying the Quantum GX2 microCT for Radiation Leakage
PerkinElmer recommends, and some local government agencies may require, an X-ray leakage safety
test be performed under the following conditions:
Every 12 months
After PerkinElmer performs maintenance or service
After any abnormal condition that could impair any of the safety systems. For example, the light
box door becomes difficult to open or close.
Conducting the X-Ray Radiation Survey
A radiation leakage test is a complicated matter requiring sensitive and expensive equipment. Some
states or localities may require special training and certification to perform the test. Contact
PerkinElmer technical support for information regarding these tests or for scheduling a PerkinElmer-
trained person to conduct the survey as part of an overall safety check.
8.2 Maintenance & Safety Checks
Daily/Weekly Safety Checks
Perform the following safety checks on a daily basis.
1. Verify that the door interlock is in good repair.
2. Verify that the key switch functions properly.
3. Verify that the "X-Ray On" indicators are functioning properly.
The following safety checks should be performed on a weekly basis.
1. All checks performed on a daily basis.
2. Inspect all screws holding the door shield and make sure that none are loose.
NOTE: There are no user-performed maintenance procedures.