13: Data Management Data Management Mode
140
The following information is listed on the display:
ID – Indicates if the event data record contains any patient information which could
uniquely identify the individual.
indicates there is; indicates there isnt.
Date and Time – Date and time the event began.
Elapsed – Duration of the event.
Paced – Total paced time for the event.
Event ID – The unique ID for the event.
Shocks – Total number of shocks delivered during the event.
Event Storage Used/ Free – The amount of space used/available in internal memory.
Number of Events Stored – The number of events that are currently stored in internal
memory.
Internal Memory Menu
From the Internal Memory Menu you can print, export, remove all patient identification data and
view data on your USB drive.
To use the Internal Memory Menu:
1 Confirm you are in Data Management.
2 Press the Smart Select knob.
3 Using the Smart Select knob, select your desired operation. See Figure 68.
4 Press the Smart Select knob to perform the task.
NOTES: Select  from the Data Management Menu to cancel an export once it begins. The
option appears in the menu after you have begun to print/export. To cancel printing, hit the Print
button.
If you change from Data Management Mode to a clinical mode while data is exporting, you are
alerted that data export is in process and asked to
 Select  to stop data export and
continue to the new mode. Select
 to continue exporting data.
If you turn the device off while exporting data, the export is stopped and the exported data might be
incomplete.
Figure 68 Internal Memory Menu







Select  to print the currently selected Event Summary.
Select
 to export the currently selected Event
Summary to the USB drive.
Select
 to export all Event Summaries currently in
Internal Memory to the USB drive.
Select
 to de-identify all Event
Summaries in Internal Memory. See “Removing All Patient
Data on page 141.
Select
 to view all Event Summaries on an
external USB drive. See Accessing Data on the USB Drive
on page 141.
Select
 to exit the menu.