Heart Rate and Arrhythmia Alarms 4: ECG Monitoring
59
Heart Rate and Arrhythmia Alarms
The Efficia DFM100 detects HR and arrhythmia alarm conditions by comparing ECG data to a set
of pre-defined criteria. An alarm can be triggered by a rate exceeding a threshold (e.g., HR >
configured limit), an abnormal rhythm (e.g., Ventricular Tachycardia) or an ectopic event (e.g., PVC
> configured limit).
HR/Arrhythmia alarms can be generated for the conditions shown in Tables 7 and 8. Once
generated, they appear as alarm messages in the HR alarm status area above the HR numeric. When
ECG alarms are off an
message appears above the HR numeric. There are both audio
and visual alerts. For more information on alarms, see “Alarms” on page 36.
NOTE: Alarm notification is configurable. See “Alarm Management” on page 38.
Table 7 HR/Arrhythmia Physiological Alarms
Alarm Message Condition Typ e of Alarm Indication
No detectable beats for four
seconds in the absence of V-Fib.
High Priority
Latching Alarm
Red Alarm message
with alarm tone
A fibrillatory wave detected for
four seconds.
Consecutive PVCs and HR
exceed configured limits.
Extreme Brady - 10 bpm below
HR low limit, capped at 30 bpm.
Extreme Tachy - Adult: 20 bpm
above HR High limit, up to 180
bpm, 200 bpm for limits
between 180-200.
Extreme Tachy - Infant/Child:
20 bpm above HR High limit,
up to 220 bpm, 240 bpm for
limits between 220-240.
For higher rates, the limit is
equal to the HR High limit.
No QRS following internal pacer
pulse.
Medium Priority
Latching Alarm
Yellow Alarm message
with alarm tone
No QRS or pacer internal pulse
detected.
(detected rate > limit)
The number of detected PVCs in
a minute exceed the limit.
Medium Priority
Latching
Configurable
Alarm
The HR exceeds the configured
HR High limit.
The HR is below the configured
HR Low limit.