EasyManua.ls Logo

Philips Efficia DFM100 - Maintenance; Overview

Philips Efficia DFM100
240 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
171
16
Maintenance
This chapter describes how to care for your Efficia DFM100 and its accessories.
Overview
Proper maintenance of the Efficia DFM100 is a simple, yet important factor in dependability.
Attending to routine maintenance is vital to keeping the Efficia DFM100 ready to respond in an
emergency.
Routine maintenance involves:
Providing power so automated tests can run (see “Automated Tests on page 172).
Observing the Ready For Use (RFU) indicator to confirm the devices readiness (see “Ready For
Use Indicator” on page 26).
Performing Operational Checks and Shift Checks (see “Operational and Shift Check” on
page 157).
Caring for batteries (see “Battery Maintenance on page 174).
Cleaning the device and accessories (see “Cleaning Instructions on page 177).
Checking expiration dates on supplies and accessories, and ordering replacements (see
Chapter 18 “Supplies and Accessories on page 195).
WARNINGS: Efficia DFM100 service should only be performed by qualified service personnel, in accordance with
the Efficia DFM100 Service Manual.
Electric shock hazards exist internally. Do not open device.