5: AED Mode Option AED View
66
AED Soft Keys
AED Mode has multiple soft keys available for use (see Figure 48):
• CPR - Pressing the
soft key initiates the configured pause period to perform CPR.
• Resume Analyzing - Pressing the
soft key initiates the AED analysis
algorithm to resume or restart analysis.
• Zero CO2 - Available if you have the CO
2
option installed and AED CO
2
monitoring
enabled. Pressing the
soft key zeros the sensor.
• Background Analysis - This key is available if your device is configured for No Shock
Advised (NSA) Monitoring and you have activated NSA Pause. Press it to begin NSA
monitoring.
Pressing the Smart Select knob brings up the main menu for AED Mode. See Figure 49.
For more information on menus, see “Menus” on page 34.
NOTE: In loud environments, use the display prompts in addition to the voice prompts.
Figure 49 AED Main Menu
Select to adjust the volume of AED
Mode voice prompts.
Select
to adjust the volume of AED
Mode alarms.
Select
to modify the ECG wave size.
See Figure 43 on page 54.
Select
to select which wave is to appear in
Wave Sector 2.
Select
to modify alarm limits.
Select
or to turn the parameter off
in AED Mode. Optioned are grayed out if they the
parameters are not configured to be on in AED
Mode.
Select
to exit the menu.