16: Maintenance Service Life
178
Printer Printhead
If the printout has light or varying print density, clean the printhead to remove any buildup of paper
residue.
To clean the printhead:
1 Open the printer door.
2 Move the paper out of the way.
3 Clean the printhead surface (top, front of the compartment) with a cotton swap dipped in
isopropyl alcohol.
4 Replace roll of paper and close door.
Side Pouches
After removing from the device, the side pouches may be cleaned by hand with mild soap and water
and air dried. Do not wash or dry by machine.
SpO
2
Sensor and Cable
Follow the manufacturer’s instructions to clean the SpO
2
sensor and cable.
CO
2
Sensor and Cable
Follow the manufacturers instructions to clean the CO
2
sensor and cable.
NBP Cuff
Follow the manufacturer’s instructions to clean the cuff.
Service Life
Life expectancies for the Efficia DFM100 therapy cable and external paddles are related to the
environment they are used in, their frequency of use and how they are cared for. Their service life is
up to three years. To maintain reliable performance and reduce the possibility of failure during
patient use, replace them every three years from the time they were initially placed into service or if
they fail inspection. Extra care must be taken when using the Efficia DFM100 in transport and
Emergency Medical Service (EMS) environments that place additional stress on the cable
connection and increase the chance for wear, impacting its useful life.
Efficia DFM100 Disposal
Prior to disposal, remove the battery. Then dispose the device and accessories in accordance with
your country’s regulations for equipment containing electronic parts.
WARNINGS: Disposing the device with the battery inserted presents an potential shock hazard.
To avoid contaminating or infecting personnel, the environment or other equipment, make sure you
disinfect and decontaminate the device and any appropriate accessories prior to disposal.