LUCAS® 2 Chest Compression System – Service Manual
3323809-001, © 2017 Physio-Control, Inc.
17
When the calibration is finished you’ll get this message:
Now the calibration result is shown in the lower right corner.
Click Store View to print and save the result.
4.5. Failed Calibration
If the value for Control exceeds 6% or if the value for Protective exceeds 3.1% the device
will make an alarm and an error code will be shown in the Control CPU window.
If this happens, mute the alarm, check that nothing interferes with the movement of the
suction cup and that the device is upright. Click POTCAL to make a new calibration.
If also the second calibration fails it might be necessary to replace the Compression Module.