9.12. Replacing the Main Body/Frame Assembly
9.12.1. Disassembling
Follow the instructions in Section 9.3 to remove the Compression Module.
Follow the instructions in Section 9.6 to remove the Controller PCB.
Follow the instructions in Section 9.7 to remove the Protective PCB.
Follow the instructions in Section 9.11 to remove the Power Inlet.
If replacing the Main Body follow the instructions in Section 9.9 to remove the
Support Legs.
Warning: Always use ESD protection when handling PCB’s!
9.12.2. Reassembling
Re-install the Power Inlet using instructions in section 9.11.2.
Install the Main Body Lid and tighten all four Torx 20 screws to 13 in-lbs / 1.5 Nm.
Re-install the Controller PCB, Protective PCB, Compression Module, and Hood.
If removed, re-install the Support Legs.
9.12.3. Test
Perform a Function Check according to Section 10.
10. Function Check
10.1. Introduction
LUCAS® 2 is an advanced system consisting of both mechanics and electronics. After each repair or
maintenance, a Function Check should be performed to establish all vital functions of the device.