LUCAS® 2 Chest Compression System – Service Manual
3323809-001, © 2017 Physio-Control, Inc.
38
8. Preventive
Maintenance
of LUCAS
NOTE: Use PIP checklist 3323774 Rev B to record PIP results during the maintenance procedure.
8.1. Yearly Maintenance Procedure
• The device is cleaned according to instructions in the IFU.
• Check that the Bellows and the Suction Cup is intact and clean, replace if necessary.
• All fabrics (Patient Straps, Stabilization Strap and the Carrying Bag) are checked with
the aspect of cleanliness and that the Velcro and buckles isn’t worn and fulfils its
function. Replace if necessary.
• The hood is opened according to instructions in Section 9.1. With the hood off,
lubricate the Compression Module according to Section 8.3.
• Connect the device to a laptop with the LUCAS 2 Program Loader Software according
to Section 4. Check the software version in the device and upgrade if possible.
• Start the device and let it run for a minute to check that no errors or warnings
occur according to section 10.
• Disconnect the USB cable and put the Hood back on according to 9.1.2. Lubricate the
Claw Lock Mechanism according to Section 8.4.
• Perform a function check according to chapter 10.
See table below:
Check/Replacement Bellows
Check/Replacement Suction Cup
Check/Replacement Patient Straps
Check/Replacement Stabilization Strap
Check/Replacement Carrying Bag
Check for Errors and Warnings
Compression Module Lubrication
Figure 8.1-1: Table –
Maintenance
procedure
* Instructions For Use
(LUCAS 2.0 and LUCAS 2.1 Art # 100666-XX)
(LUCAS 2.2 Art # 100901-XX)