LUCAS® Chest Compression System Service Manual
3323809-001, © 2017 Physio-Control, Inc.
59
LUCAS 2.2
Re-install the Hood.
9.8.3. Test
Perform a Function Check according to Section 10.
When the device is in active mode let it run, after approximately 10 minutes the fan should
start. Now set the device in adjust mode, the fan should continue running. Check that there is
no unusual sound from the fan. When the temperature inside the hood has decreased to
30°C the fan should stop. Continue with the Function Check.