LUCAS® 2 Chest Compression System – Service Manual
3323809-001, © 2017 Physio-Control, Inc.
43
Remove the two mounting screws to remove the Hood
Bracket if replacement is needed Torx 20 (T20).
9.1.2. Reassembling
Install the Hood Bracket if removed and tighten the two
Torx 20 screws to 13 in-lbs / 1.5 Nm
Re-connect the three cables.
Carefully put the hood back in place, make sure not to damage any cables.
Lock the four screws with thread lock fluid and tighten to 5 in-lbs / 0.6 Nm Torque wrench
PH1.
9.1.3. Test
Perform a Function Check according to Section 10.