LUCAS® 2 Chest Compression System Service Manual
3323809-001, © 2017 Physio-Control, Inc.
24
LUCAS 2.0 and LUCAS 2.1
5.2-2: Table
Malfunction
alarms LUCAS 2.0 and LUCAS 2.1
5.3. Error Codes
In the software, LUCAS 2 Program Loader, it is possible to read error codes from the device.
Below tables show these codes and an explanation on how to find the cause and point out which
module to replace.
5.3.1. Error Codes from the Control CPU
Error Code
Explanation
Probable cause
Module to check /
replace
T1_C_EE
No=0x10
Control EE-Prom
CheckSum data
error
CPU fault
Replace Control board
T1_START_CO
M
No=0x13
No Communication
between Protective
and/or Battery-
Charger CPU's at
start
I2C
communication
faults.
Halted Protective or
Charger CPU.
Replace cable between
boards or Protective
board.
T1_PB_FLASH
No=0x14
Incompatible
programs in
Protective board
Change of boards.
Reprogram device.