LUCAS® 2 Chest Compression System Service Manual
3323809-001, © 2017 Physio-Control, Inc.
41
9. Replacing
Parts/Modules
in LUCAS
IN THIS SECTION IT IS EXPLAINED HOW TO REPLACE SOME OF THE MOST COMMON PARTS/MODULES.
HOW TO APPLY THREAD LOCK FLUID
The normal amount of Thread Lock Fluid should be corresponding to the diameter of the
screw and the Thread Lock Fluid should be applied at the lower end of the screw, see picture
below:
Thread Lock Fluid
This general description shall apply on all use of Thread Lock Fluid on LUCAS according to
this manual.
WARNING: TAKE OFF THE BATTERY BEFORE OPENING THE DEVICE!
Do not connect the battery to hood cover without the hood holder service tool installed. The
weight of the battery in a hood cover that isn’t installed on the device can damage the wires
or connections
Use caution when examining or operating the device without its covers.
9.1. Replacing the Hood and Hood Bracket
9.1.1. Disassembling
Remove all four screws for the hood PH1. Set aside screws for reuse during assembly.