LUCAS® Chest Compression System Service Manual
3323809-001, © 2017 Physio-Control, Inc.
69
9.11. Replacing the Power Inlet
9.11.1. Disassembling
Follow the instructions in Section 9.3 to remove the Compression Module.
Follow the instructions in Section 9.6 to remove the Controller PCB
Follow the instructions in Section 9.7 to remove the Protective PCB
Warning: Always use ESD protection when handling PCB’s!
Remove the four Torx 20 screws to remove the Main Body Lid.
Use the extraction tool to remove the wires from the Power Inlet.
With the wires removed you can now remove the nut and Power Inlet.
9.11.2. Reassembling