LUCAS® Chest Compression System Service Manual
3323809-001, © 2017 Physio-Control, Inc.
56
9.6.2. Reassembling
Put the new PCB in place and fasten loosely with the screw together with the Bottom Plate
Torx 25 (T25).
Re-install the Compression Module and connect all cables. Connect the
cables for the Hood, but do not fasten it.
Re-install the Bellows.
Re-install the Suction Cup.
9.6.3. Programming
Connect a laptop and program the Control PCB according to Section 4.
Tighten the Hood.
9.6.4. Test
Perform a Function Check according to Section 10.
9.7. Replacing the Protective PCB
9.7.1. Disassembling
Follow the instructions in Section 9.3 to take out the Compression Module.
Remove the remaining screw from beneath and lift out the PCB carefully.
Warning: Always use ESD protection when handling PCB’s!
9.7.2. Reassembling
Put the new PCB in place and fasten loosely with the screw together with the Bottom Plate
Torx 25 (T25).
Re-install the Compression Module and connect all cables. Connect the cables
for the Hood, but do not fasten it.
Re-install the Bellows.
Re-install the Suction Cup.
9.7.3. Programming
Connect a laptop and program the Protective PCB according to chapter 4.
Check that Device ID, Time and Date is correct.
Tighten the Hood.