LUCAS® Chest Compression System Service Manual
3323809-001, © 2017 Physio-Control, Inc.
60
9.9. Replacing the Support Leg
9.9.1. Disassembling
LUCAS 2 devices with S/N up to 3012 4632 (LUCAS 2.0 and LUCAS 2.1) have
Shaft Seals assembled. On LUCAS 2 devices with a S/N greater than 3012 4632,
the Shaft Seals are replaced with a Snap Ring and a Bumper with integrated shaft
seal.
Remove the Patient Strap. Remove the Bumper (if installed)
Remove the Shaft Seals by bending with a Flat Screwdriver. If the LUCAS 2
device has Snap Rings installed instead of Shaft Seals, remove Snap Rings.
Remove the Patient Strap Holders.