LUCAS® Chest Compression System – Service Manual
3323809-001, © 2017 Physio-Control, Inc.
67
Slide the Bumper up over the support leg (note the direction)
Thread the Bumper over the shaft seals and the patient strap holder on both sides.
Note: There is only Shaft Seals on older LUCAS 2
Mount back the Patient Strap.
If the replaced Support Leg had a Type Label with serial number, contact Physio-Control to
obtain a new label.
9.9.3. Test
Do the following tests to verify that the repair/replacement has been performed correctly:
Check the locking function by locking and unlocking against the Back plate.
Check the function of the torsion spring by folding and unfolding the Support legs.