Note that depending on the selection for Automatic I-Term calculation, the Servo-
Loop I-Term value is changed too automatically when you change the Notch
Frequency 1 value (for details, see p. 134).
b) If you have measured a second resonance peak, enter the frequency value of the
second resonance peak in the Notch Frequency 2 field (in Hz). If the second notch
filter is deactivated (rejection value = 1), change the rejection value to 0.05 in the
Notch Rejection 2 field.
8.
Save or discard the new settings in the Parameter Settings panel:
If you want to keep the new settings, save them to the nonvolatile memory of the
E-727 by clicking the Save as Default (EEPROM) button.
If you want to discard the new settings and reset the parameter values to their
defaults (i.e. to their values from nonvolatile memory), click the Reset All to
Default button.