User Manual
E727T0005, valid for E-727
BRO, 2019-06-28
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 180 / 240
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Response to this command:
<Arg3>"="<Val3>SPLF
<Arg1>"="<Val1>SPLF
<Arg2>"="<Val2>LF
Example 5:
Send: TSP?SP2SP1LF
Receive: 2=-1158.4405SPLF
1=+0000.0000LF
Limitations for GCS Commands
More than one command mnemonic per line is not allowed.
The number of characters per line is limited to 256 byte (1 character = 1 byte). This means that the
number of arguments following a command mnemonic is limited to 32.
Example:
If you send
TWS 1 100 1 1 200 1 1 300 1 1 400 1 1 500 1 1 600 1 1 700 1 1 800 1
1 900 1 1 1000 1 1 1100 1
the controller will return error 24 ("Incorrect number of parameters") when you ask with the ERR?
command afterwards because the number of arguments is 33.