E727T0005, valid for E-727
BRO, 2019-06-28
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 53 / 240
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Note: The input and output signal channels of the E-727 are allocated to the logical axes via
matrices (input matrix: parameters 0x07000500 to 0x07000506; output matrix: parameters
0x09000000 to 0x09000003). Depending on the connected stage type, an axis may be driven
by more than one piezo actuator and measured by more than one sensor.
Creating Backup Files for Controller Parameters
INFORMATION
The properties of the E-727 and the connected stage(s) are stored in the E-727 as parameter
values.
Create a backup copy on the PC before changing the parameter values of the E-727. You can
then restore the original settings at any time.
Create an additional backup copy with a new file name each time after you optimize the
parameter values.
To save the parameter values and to load them back to the E-727, use the Device Parameter
Configuration window provided by PIMikroMove.
Proceed as follows to create a parameter file:
1. In the main window of PIMikroMove, open the Device Parameter Configuration window via
the E-727… > Parameter Configuration menu item.