Planmeca ProMax 3D Max with ProTouch 23
3D PATIENT EXPOSURE
User’s Manual
The preset exposure values are shown in the following
tables.
Factory presets for image resolution Low dose
in programs Nose and Sinus
PATIENT SIZE kV VALUE mA VALUE
Child (XS) 96 2
Small adult (S) 96 2.5
Medium-sized adult (M) 96 3.2
Large adult (L) 96 4
Factory presets for image resolution Low dose
in other programs
PATIENT SIZE kV VALUE mA VALUE
Child (XS) 96 4
Small adult (S) 96 5
Medium-sized adult (M) 96 6.3
Large adult (L) 96 8
Factory presets for image resolution Normal
in programs Nose and Sinus
PATIENT SIZE kV VALUE mA VALUE
Child (XS) 96 4
Small adult (S) 96 5
Medium-sized adult (M) 96 6.3
Large adult (L) 96 8
Factory presets for image resolutions Normal, HD, Hi Res and Endo
in other programs
PATIENT SIZE kV VALUE mA VALUE
Child (XS) 96 6.3
Small adult (S) 96 8
Medium-sized adult (M) 96 10
Large adult (L) 96 12.5