CLEANING
50 Planmeca ProMax 3D Max with ProTouch
User’s Manual
12 CLEANING
NOTE Disconnect the X-ray unit from mains before cleaning.
NOTE Do not use cleaning solutions in aerosol or spray form
directly on unit surfaces.
X-ray unit
Sensors
13 SERVICE
To guarantee patient and user safety and to ensure
consistent image quality the X-ray unit must be checked
and recalibrated by a qualified Planmeca service
technician once a year or after every 10 000 exposures if
this is sooner.
Autoclave
up to 135°C
Wipe with
soft cloth using
alcohol based
cleaning solution
Wipe with
soft cloth using
mild cleaning
solution
3D head supports
(incl. support bars)
X
Chin cup XX
Patient handles X
Other surfaces
(incl. touch screen)
X
Wipe with soft cloth
(NO CLEANING
SOLUTION)
Compressed air
3D sensor X
ProFace sensor;
laser windows in the middle
X
ProFace sensor;
other surfaces (incl. glass
windows on both sides)
X