Planmeca ProMax 3D Max with ProTouch 53
HELP MESSAGES
User’s Manual
H151 Line voltage The line voltage was too low during
exposure.
Exposure was interrupted.
Contact your service technician
for help.
H152 The line voltage is too low. Exposure is not possible.
Contact your service technician
for help.
H161 Temperature The temperature of the tube head is too
high.
Wait for a few minutes for the
tube head to cool down.
H162 The temperature of the lift motor is too
high.
Wait for a few minutes for the lift
motor to cool down.
H163 The temperature of the power supply
unit (PSU) is too high.
Wait for a few minutes for the
power supply unit (PSU) to cool
down.
H165 The temperature of the tube head is too
high for the selected exposure values.
Wait for a few minutes for the
tube head to cool down.
H166 The maximum tube head energy level
was exceeded.
Wait for a few minutes for the
tube head to cool down or use
lower exposure values.
H177 User related
messages
Exposure is not possible with these
settings.
Change the image volume
settings.
H178 Exposure is not possible with these
settings.
Change the settings.
H181 The imaging process was cancelled in
Planmeca Romexis.
H182 Timeout in image data transmission. Exposure was interrupted.
Contact your service technician
for help.
H187 Problem during image data
transmission.
Exposure was interrupted.
Contact your service technician
for help.
H189 The screen was touched during
exposure.
Exposure was interrupted.
H191 The y coordinate is out of limits. Set the y coordinate between
51-91 mm.
H192 It is not possible to enable radiation or
PC communication when demo licenses
are switched on.
First switch off demo licenses,
then enable radiation or PC
communication.
H195 The 3D Model programs must not be
used for patient imaging.
Use the 3D Model programs for
taking exposures of impressions
or plaster casts only.