Section 2 Specifications
2.1 Technical specifications
Specifications are subject to change without notice.
Sample
Number of channels 1 - 6
Measurement cycle < 10 min / channel
Sample pressure 0.2 to 6 bar (3 to 87 psi)
Temperature 5 - 50°C (41 - 122°F)
Sample flow
Minimum 5L / hour
Maximum 30L / hour
Connections
Sample input
Simple fittings for 6 mm O.D. tubing or ¼" O.D. in PE-low density.
¼" OD in PHED-PTFE-SS as option
Sample output Barbed stem for 12 mm (½" I.D.) hose
Conditions of use
Ambient temperature 5 to + 45°C (41 to 113°F)
Power supply voltage fluctuation ± 10%
Overvoltage category 2 (according to standard EN 61010-1)
Degree of pollution 2 (according to CEI 664)
Altitude 2000 m
Relative humidity 10 to 80%
Measurement category Cat II, Class 1 (overvoltage < 1500V)
Weight
Panel analyzer without canister 13 kg
Cabinet analyzer without canister 65 kg
Full canisters 8 kg
Standards
EMC EN 61326-1 (1997) & A1 (1998) & A2 (2001) & A3 (2004)
European safety standards IEC 61010-1 (2001) & EN 61010-1 (2001)
UL & CSA agreement
Certified to UL & CSA safety standards, with cETLus mark
(Control Number: 3164002)
UL & CSA standards
Safety of Electrical Equipment for Measurement, Control and Laboratory use;
UL61010-1, Issued: 2004/07/12, Ed.2 Rev: 2005/07/22 and CAN/CSA C22.2
No. 61010-1, Issued: 2004/07/12
Analysis
Value measured Dissolved SiO
2
Cycle time Approximately 10 minutes per channel
Measurement range (2 versions) 0 - 1000 ppb 0 - 5000 ppb
Repeatability ± 2% or ± 0.5 ppb ± 2% or ± 2 ppb
Detection limit 0.5 ppb 2 ppb