EasyManua.ls Logo

Polymetron 9210 - Page 77

Polymetron 9210
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
75
Default Configuration
mA OUTPUTS / AFFECT / OUTPUT 2
Affect Channel 2 Channel 1 - 6
Type 0-20 mA 0/20 mA / 4/20 mA
Mode Linear Linear / dual
Low 0 ppb
Middle 100 ppb
Up 1000 ppb
mA OUTPUTS / AFFECT / OUTPUT 3
Affect Channel 3 Channel 1 - 6
Type 0-20 mA 0/20 mA / 4/20 mA
Mode Linear Linear / dual
Low 0 ppb
Middle 100 ppb
Up 1000 ppb
mA OUTPUTS / AFFECT / OUTPUT 4
Affect Channel 4 Channel 1 - 6
Type 0-20 mA 0/20 mA / 4/20 mA
Mode Linear Linear / dual
Low 0 ppb
Middle 100 ppb
Up 1000 ppb
mA OUTPUTS / AFFECT / OUTPUT 5
Affect Channel 5 Channel 1 - 6
Type 0-20 mA 0/20 mA / 4/20 mA
Mode Linear Linear / dual
Low 0 ppb
Middle 100 ppb
Up 1000 ppb
mA OUTPUTS / AFFECT / OUTPUT 6
Affect Channel 6 Channel 1 - 6
Type 0-20 mA 0/20 mA / 4/20 mA
Mode Linear Linear / dual
Low 0 ppb
Middle 100 ppb
Up 1000 ppb
mA OUTPUTS / SPECIAL PROG / MAINTENANCE
Mode Last Last / Preset / Live
Preset 0 mA
mA OUTPUTS / SPECIAL PROG / CALIBRATION
Mode Last Last / Preset / Live
Preset 0 mA
mA OUTPUTS / SPECIAL PROG / GRAB SAMPLE
Mode Last Last / Preset / Live
Preset 0 mA
User configuration

Table of Contents